Bubbler Firmware Version 1.7.379 released.
This firmware update is to address an issue reported with some EPDM valves that were found to be sticking due to sitting on the shelf over the winter. The issue was highlighted by a Pump-Rate-Low error message being reported by the Bubbler. A longer pulse width of 15mS will be applied to Bubblers that have an EPDM valve installed.
Update Instructions.
(1) Power down the Bubbler by removing both green connectors (battery and SDI)
(2) Open the Bubbler and remove the SD card. This is located on the left hand side pcb.
(3) Download the firmware update and save to the SD card.
(4) Re-insert SD card and close Bubbler.
(5) Power up
(6) Diagnostics menu => Update.